11
A-MIP 210 V1.06
In case of breakdown
• White indicators do not light up at the aivia energizing.
1. Check the wiring polarity.
2. Check the power supply cable connection.
3. Check the voltage at the connector block.
4. Check the insertion of the connector block.
5. Contact your partner.
• White indicators do not blink in night operation.
1. Check the aivia energizing.
2. Contact your partner.
• Red indicators do not blink when the door is being opened.
1. Check the aivia energizing.
2. Contact your partner.
• Siren does not works when the door is opened.
1. Check for the siren option on the Avia identification plate.
2. Check the aivia energizing.
3. Contact your partner.
• I do not know my maintenance/user code.
.
1. Shut down the Aivia power supply.
2. Open the Aivia.
3. Energize the Aivia, the white indicators will be lighted.
4. Set the User/Maintenance code, as described in the « Setting up the Aivia » section.
5. Close the Aivia.
• Red indicators blink in a repeated way.
• Aivia motorization doesn’t work.
Withdraw the cap off the aivia door, and push back the door blocking axle with a stem .
Number of blinkings Cause Solution
1
Failing memory Contact your partner.
2
Parameters lost Set-up the Aivia again
3
Temperature sensor breakdown Contact your partner
4
Alert temperature reached Check AED operates correctly, if needed, contact your partner.
Continuous
Aivia has been opened Check AED operates correctly, and presence of all the accesso-
ries, if needed, contact your partner.
Cap