EasyManua.ls Logo

Alcoa 200 - Troubleshooting; Tool Malfunction Causes and Remedies

Alcoa 200
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
200 Series Tooling Alcoa Fastening Systems
16
T
T
ROUBLESHOOTING
ROUBLESHOOTING
TOOL MALFUNCTION CAUSE AND
TOOL MALFUNCTION CAUSE AND
REMEDY
REMEDY
A. Air supply pressure low. Air pressure of 90-100 psi
required.
B. Air in hydraulic system. Refill and bleed tool per
instructions.
C. Check nose assembly for completeness and prop-
er installation per instructions.
D. Shift valve adjustment set too high (13-7). Readjust
counter-clockwise per instructions.
E. Shift valve adjustment set too low (13-7). Re adjust
clockwise per instructions.
F. Relief valve (13-8) not sealing. Check 3/16 inch
ball (2-47) and seat (2-3) for foreign particles.
Clean if dirty. Reseat ball by tapping against brass
seat. Replace ball or seat if damaged.
G. Check valve (13-9) for leakage. Clean and reseat
ball (2-51) by tapping. If ball or seat (2-2A) are
damaged, replace.
H. Bleed screw (2-40) not backed out to stop (13-11).
Balance valve (2-33) stuck in open position or for-
eign material between valve and seat. Free valve
and remove any foreign particles. If valve or seat
(2-4) are damaged, replace.
I. Check valve (13-10) for leakage. Clean valve and
reseat ball (2-51) by tapping. If ball or seal (2-2B)
are damaged, replace.
J. Shift sleeve (4-27) loose (13-12). Tighten with
Wrench 100878.
K. Shift valve seat (4-13) not sealing (13-13). Replace
seat if foreign particles are imbedded in seal, or if
seat is damaged.
L. Shift plunger (4-10) sticky in sleeve (4-7). Remove
plunger from sleeve. If plunger or sleeve bore is
scored, replace (13-13).
M. Extreme temperature variation. Adjust shift valve
(13-7) as necessary.
CAUTION: Make certain that Valve compo-
nents are not magnetized.