EasyManua.ls Logo

Allen-Bradley Guardmaster - Check Voltage-Free Contacts

Allen-Bradley Guardmaster
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rockwell Automation Publication 440R-UM013E-EN-P - March 2017 67
Troubleshooting Chapter 11
Table 14 shows the voltage levels that are viewed on an oscilloscope versus a
digital multimeter.
Check Voltage-free Contacts
Safety devices (for example; interlock switches, E-stops, or cable pull switches)
with voltage-free contacts must be connected to the pulse testing outputs. You
can use a digital multimeter to measure the input levels.
Check CI Safety Relay
1. With the device contacts open, measure the voltage at the pulse testing
outputs with a digital multimeter, as shown in Figure 60
. The voltage
must be 18…19V on both pulse testing outputs of the CI safety relay.
Figure 60 - Typical Voltage Measurements of the CI Safety Relay
IMPORTANT The following factors affect the value that is measured at the relay inputs:
•Voltage-free contacts
Pulse testing waveforms
Capacitance
•Length of wire
Contact resistance
Channel sequence
Table 14 - ON/OFF Voltage
Measurement Device Turn ON Voltage Turn OFF Voltage
Oscilloscope 11V 5V
Digital Multimeter 6…8V 3…4V
Pulse Testing
Outputs
S11
S21
S12
19
19
S22
A2
Volts
DMM
Volts
DMM

Table of Contents

Related product manuals