which is inside the equipment. The adjustment can have a tolerance of 20%. Verify that the output pressure
does not exceed 2 bars. The nominal value is around 1.5 bars.
LIST OF THE ERROR CODES
In order to facilitate the technical assistance, an error code for each anomaly is memorized inside the
E
2
PROM; thanks to a particular procedure, which can be selectable exclusively in the calibration mode, it is
possible to visualize the last 32 errors that occurred to the unit. It is also possible to cancel this list, in order
to avoid repeated checks when the equipment is switched on again.
HOW READING THE LIST OF THE ERRORS
To visualize the list of the errors, it is necessary to enter into the calibration mode, by pressing at the same
time the three buttons: CAL, selection of the bipolar coagulation mode, and increase of the monopolar
coagulation power. After the appearance of the signal message about the calibration status on the display
(CAL on), press the CAL button again, and, with the blinking displays, select 0d into the monopolar Cut
section and a number from 00 to 20 (from 0 to 20 in the hexadecimal scale corresponds to 32 positions in
the decimal scale) into the monopolar coagulation section (the latter represents the number of the
memorized error; the last error that has occurred is indicated by the signal on the Cut display that will be
FF). By pressing the monopolar Cut activation pedal-switch, the error code about the position selected
before with the monopolar Coagulation section will appear.
HOW DELETING THE LIST OF THE MEMORIZED ERRORS
Always in the calibration mode, press the CAL button and, with the blinking displays, select 0E into the
monopolar Cut section; by pressing the monopolar Cut activation pedal-switch, the display will show the
code E2P CLr, to inform the operator that the memories for the errors have been deleted. By repeating the
reading sequence of the errors (see the previous paragraph), all the memories will contain FF.