2.0 Safety instructions
Read this manual carefully before using the
instrument!
In order to avoid any personal injuries and/or damage to
the instrument all safety instructions and warnings must be
observed.
Disposal – WEEE Directive
These calibrators contain Electrical and Electronic circuits
and must be recycled or disposed of properly (in
accordance with the WEEE Directive 2002/96/EC).
Warning……
About the use:
• The calibrator must not be used for any purposes other
than those described in this manual.
• The calibrator has been designed for interior use only
and should not be used in hazardous areas, where
vapour or gas leaks, etc. may constitute a danger of
explosion.
• The calibrator must be kept free within an area of 20
cm on all sides and 1 metre above the calibrator.
• Never use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
About the front panel:
• The connections used to test thermostats must NEVER
be connected to a voltage source.
• Thermostats must not be connected to any other
voltage source during a test.