3.0 Operating the calibrator
3.1 Before use
Warning
The calibrator must not be used for any purposes other
than those described in this manual.
The calibrator has been designed for interior use only
and should not be used in hazardous areas, where
vapour or gas leaks, etc. may constitute a danger of
explosion.
The connections used to test thermostats (fig. 1 pos. 8
and 9) must NEVER be connected to a voltage source.
Thermostats must not be connected to any other
voltage source during a test.
NEVER use heat transfer fluids such as silicone, oil,
paste, etc.
These fluids may penetrate the calibrator and cause
damage or create poisonous fumes.
The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator.
Caution – Hot surface
This symbol is engraved in the grid plate.
Do not touch the grid plate, the well or the insertion
tube when the calibrator is heating up – they may be
very hot.
Do not touch the handle of the calibrator during use – it
may be very hot.

Other manuals for Ametek JOFRA ITC-320 A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ITC-320 A and is the answer not in the manual?

Ametek JOFRA ITC-320 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ITC-320 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals