EasyManua.ls Logo

Ametek mocon Dansensor ISM-3

Ametek mocon Dansensor ISM-3
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 User Manual EN Dansensor® ISM-3
P/N 200114-I
09/2021
If the ISM-3 is equipped with an internal pump, the gas is sucked from measure point at
atmospheric pressure. When calibrating, the gas is sucked from a gas overflow, to which
approx. 0.5 l/min must be supplied.
Fig. 6. Flow diagram of ISM-3 with pump. Gas overflow for calibration use is shown
Operation and connection
Operation of ISM-3:
Display 4 digit red LED display
Control 4 front keys. Key function is activated when key is released.
Measuring accuracy Better than +/- 1% of the measured value
+/- 1 digit in calibrated measuring range
Measuring range 20.9% - 1ppm (standard) *
* For measuring in other ranges device must be recalibrated - see "Calibration" on page 19 for details
Possible connection to the ISM-3:
Gas supply Gas inlet on the back
Relay Max. 48V, 1A. (COMMON, N.O. and N.C.)
Current output Programmable 0-20 or 4-20 mA, with user-defined scale
(e.g. 0-1%, 10-20.9% or 0-100ppm O
2
).
Voltage output (Option) Programmable 0-10 volt or 2-10 volt with user defined scale
Measure input 10-32VDC external pump control. Consumption 10mA.
RS232C (Option) For serial communication with a PC (for internal use only)
All electrical inputs and outputs, except for the RS232C port, is galvanically separated from
the internal electronics by means of optocouplers and/or a relay.
(for flow indication)
Restriction hose
Gas out
125 ml/min
Pump
Gas in
0,5 l/min
Hoses:
ID = min. 3 mm
L = min. 0.5 m
O
2
sensor
Pressure
sensor

Related product manuals