20
15120-101 Rev. D
Guidance Tables (continued)
Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
TheXcel255isintendedforuseinelectromagneticenvironmentspeciedbelow.Thecustomer
or user of the Xcel 255 should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
- Guidance
ESD
EN/IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete or ceramic
tile.Ifoorsaresynthetic,theR/Hshouldbeat
least 30%.
EFT
EN/IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
±1kV I/Os
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
EN/IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage
Dips/Dropout
EN/IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
Note 1
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Xcel 255
requires continued operation
during power mains interruptions, it is recom-
mended that the Xcel 255 be powered from an
uninterruptible power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
3A/m 3A/m
Powerfrequencymagneticeldsshouldbethat
of a typical commercial or hospital environment.
Note 1: During a 5 second loss of power, the Xcel 255 powers off, but recovers once power is restored; meeting the
manufacturers performance requirements.