Calibration and Troubleshooting 
 
7-8  M370430-01  Rev F 
Offset Calibration 
Offset calibration of the power supply provides the best accuracy in 
low-range current. 
To perform offset calibration: 
1.  After performing gain calibration, set the current to 10% from 
the nominal. 
2.  Read the current value on the ammeter display. 
3.  Type SCPI command with the current value as parameter 
“data”. 
For example, you would type : CAL : OUT P: CURR 0.50 if you 
had an XG 150-5.6 calibrating with the current set point set to 0.56 
A and the actual current measured was 0.50A. 
After performing offset calibration, it is highly recommended that you 
repeat gain calibration. 
Important: For best results, both calibrations may be repeated several times. 
To maintain the high current accuracy of the XG family of power supplies, user 
current calibration should only be done with a high accuracy calibrated shunt 
resistor of at least 0.025% tolerance. 
SCPI Command 
[:]CALibration[<channel>]:OUTPut:CURRent{<ammeter 
reading in amps>}