Local Operation
3-36 M370430-01 Rev F
4. Turn to display or .
5. Press to commit the selected setting.
Important: This single front panel operation affects both power on and
output enable. However, when using SCPI, there is a separate command
for each.
The SCPI command to query or enable/disable the 0 Voltage
Output Preset at power-on is:
[:]SYSTem[<channel>]:PON:VOLTage {?|ON|OFF}
The SCPI command to query or enable/disable the 0 Voltage
Output Preset upon output enable is:
[:]SYSTem[<channel>] :PON:ENAble:VOLTage {? |ON|OFF}
Programming Current Output Preset
To Enable/Disable automatic reset to 0 Amps:
1. Turn the rotary Adjust/Enter control to PRT and press.
2. Turn again until is displayed.
3. Press to display .
4. Turn to display or .
5. Press to commit the selected setting.
Important: This single front panel operation affects both power on and
output enable. However, when using SCPI, there is a separate command
for each.
The SCPI command to query or enable/disable the 0 Current
Output Preset at power-on is:
[:]SYSTem[<channel>]:PON:CURRent {?|ON|OFF}
The SCPI command to query or enable/disable the 0 Current
Output Preset upon output enable is:
[:]SYSTem[<channel>] :PON:ENAble:CURRent {? |ON|OFF}