14 SPECTRO ARCOS FHX22 —
The instrument is equipped with a peristaltic pump with rotating parts.
Ensure that you and other objects keep out of harm's way while the pump is
running.
Compliance with the relevant precautionary procedures specific to the liquid
and application in question is required when conveying hazardous
substances.
Warning! Risk of suffocation!
The instrument is operated with argon. Argon is heavier than air. In case of
leakage, argon can accumulate in enclosed spaces and especially on the
floor or lower-lying areas.
Ensure that adequate ventilation is provided. If necessary, install oxygen
sensors.
Warning! Optical radiation!
Plasma emits UV radiation. A protective pane is used to shield from
radiation.
Operate the instrument only with the protective pane in place and the door
closed.
-frequency emitter is used to operate the ICP, dependent on the type of
model. The electromagnetic emission is below the legally specified limit values.
Please point out to your employees that an electromagnetic emitter is used,
by attaching the prohibition sign "Danger! Persons with pacemakers not
permitted"!
Information
In order to ensure the availability and operational safety of the equipment, we
recommend having the system checked for correct operation and performance at
least once a year by specialized technicians. Addresses of service organizations
authorized by SPECTRO can be found at www.spectro.com.