3 Technical data
Description Value
Instrument height 1068 mm; 42.1"
Instrument width 1582 mm; 62.3"
Instrument depth 756 mm; 29.8"
Instrument weight approx. 240 kg; ~ 530 lbs
Footprint 1065 x 646 mm; 41.9 x 25.4“
Power supply 230 V AC ± 5%, 50/60 Hz
Phase + neutral conductor + PE; the
is equipped with a CEE 32A plug.
Instrument power consumption 4.0 kVA
Fuse protection of instrument 30 to 32 A (slow-blow)
Sound pressure level 57 dBA
Stability
(according to DIN EN 61010-1)
compliant if change of position <10°
The instrument is front-heavy.
Storage conditions 2 to 40°C; 36 to 104°F
Operating range (permissible ambient
temperature)
15 to 35 °C; 59 to 95°F
Recommended temperature range to
achieve the specified performance
18 to 25°C; 64 to 77°F, maximum temperature
gradient ∆T = 3°C/h (5.4°F/h)
Atmospheric humidity < 80% rel. (non-condensing)
Air quality Free of corrosive components and high
concentrations of dust