EasyManua.ls Logo

Ametek SPECTRO ARCOS FHX22 - Exhaust Air (to the Outside)

Ametek SPECTRO ARCOS FHX22
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Transportation/Setting-
up
SPECTR
O ARCOS FHX2218.07.2019
21
4.7 Exhaust air (to the outside)
Warning of toxic substances!
Vapors may be generated from the sample material. These may be toxic
and/or corrosive.
In these cases, provide protective measures, e.g. breathing apparatus
The sample material may be toxic and/or corrosive.
Ensure that the external exhaust system functions properly.
Exhaust air nozzles are located at the top of the instrument.
The corrosion-free exhaust chimneys must be connected with the laboratory
exhaust system using a non-flammable, flexible and corrosion-resistant hose (ID
100 mm). The fan must be resistant to chemicals and designed to operate at
temperatures up to 80°C (176°F). The exact position of the exhaust nozzles is
shown in the figure below.
All installation must be performed in a professional manner and in accordance with
applicable engineering standards in order to prevent risk to life and limb and avoid
damage to material property.
Fig. 7
Extraction capacity at the nozzle
1 (rear)
Generator
2 (front)
Plasma torch box housing
250300 m³/h; 150177 CFM Depending on the ambient temperature
1825°C = 200250 m³/h; 118150 CFM
> 25°C = 250300 m³/h; 150177 CFM
1
2

Table of Contents