16 SPECTRO ARCOS FHX22 —
Description Value
Argon gas connection
Inlet pressure during operation 6 to 8 bar (87 to 116 psi)
It is imperative not to exceed a maximum Ar
inlet pressure of 8.0 bar (116 psi).
Quality ≥ 4.6 (99.996 %)
Consumption ≤ 25 l/min
Oxygen supply for auxiliary gas (optional; for organic applications only)
Inlet pressure during operation 4.0 bar (58 psi)
Quality ≥ 4.5 (99.995 %)
Flow rate The flow rate is application-dependent.
Extraction (continuous)
Plasma torch box Dependent on ambient temperature
18 to 25 °C = 200 to 250 m³/h; 118 to 150 CFM
> 25°C = 250 to 300 m³/h; 150 to 177 CFM
Generator 250 to 300 m³/h; 150 to 177 CFM