EasyManua.ls Logo

Amico Alert-2 - Page 16

Amico Alert-2
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16 Amico Pipeline
When you have completed step #7, the display module
will automatically go into a “RESET mode. This will store the
data that you had entered.
PSI / kPa / BAR selection
Factory Default - PSI
For PSI mode, set the switch #4 to the ON position. The LED PSI indicator located next to the GAS
pressure reading will illuminate.
For kPa mode, set the switch #4 to the OFF position and switch #9 to the ON position. The LED kPa
indicator located next to the GAS pressure reading will illuminate.
For BAR set the switch #4 to the OFF and the switch #9 to the OFF position. The LED kPa indicator
located next to the GAS pressure reading will illuminate. (There is no separate indicator for BAR).
VACUUM ONLY
Vacuum alarm set-point adjustment
Factory Default:
High = 30”Hg, Low = 12”Hg
Repeat time = 30 min.
During programming, the Trend Bar will Flash!
1. Set switch #6, #7 and #8 to the ON position.
2. The LED will display (HI-), followed by the current set point. Indicating the system is ready to
accept a new High set point. Do not adjust this set point since the High set point is not used.
3. Set switch #7 to the OFF position.
4. The LED will display (LO-), followed by the current set point. Indicating the system is ready
to accept a new Low set point. Adjust set point, using the “UP and “DOWN” push buttons, to
the desired value.
5. Set switch #8 to the OFF position.
6. The LED will display (I-I-), followed by the current set point. Indicating the system is ready to
accept a new Repeat time set point. Adjust set point using the “UP and “DOWN”push buttons,
to the desired value. [(Display dd=Disabled) Range from 1 to 60 Minutes]
7. Set switch #6 to the OFF position.
When you have completed step #7, the display module will automatically go into a “RESET mode.
This will store the data that you had entered.
PSI
KPA
BAR
Alert - 2 Series
Page: 17
THE AREA/2 IN 1 DISPLAY MODULE
A dip switch is located on the back of the display module
which is used to identify the gas of the display module. The
dip-switch contains ten switch settings.
PRESSURE ONLY
Factory Default:
High = 60 Psi, Low = 40 Psi
Repeat time = 30 min.
HIGH PRESSURE/NITROGEN
Factory Default:
High = 195 Psi, Low = 140 Psi
Repeat time = 30 min.
During Programming the
“Trend Bar” will Flash!
1. Set switch #6, #7 and #8 to the ON position.
2. The LED will display (
HI-HI-
HI-HI-
HI-
), followed by the current set point.
Indicating the system is ready to accept a new High set point.
Adjust set point, using the “UP” and “DOWN” push buttons, to the
desired value.
3. Set switch #7 to the OFF position.
4. The LED will display (
LO-LO-
LO-LO-
LO-
), followed by the current set point.
Indicating the system is ready to accept a new Low set point.
Adjust set point, using the “UP” and “DOWN” push buttons, to the
desired value.
5. Set switch #8 to the OFF position.
6. The LED will display (
I-I-I-I-
I-I-I-I-
I-I-
), followed by the current set point.
Indicating the system is ready to accept a new Repeat time set
point. Adjust set point using the “UP” and “DOWN” push buttons,
to the desired value. [(Display dd=Disabled) Range from 1 to 60
Minutes]
7. Set switch #6 to the
OFF position.
2 in 1 Alarm
Board
Area Alarm
Board
When you have completed
step #7, the display module
will automatically go into a
“RESET” mode. This will store
the data that you had entered.

Table of Contents

Other manuals for Amico Alert-2