www.amico.com 53
Appendix P
Technical Specication
Supply Voltage: 115 to 220 VAC - 50 to 60 Hz
Current Draw: 1 Amp.Max.
Fuse (1/4 * 1-1/4): Fast Blow 1 Amp.
Cable requirement:
Area Display Module to Remote Sensor:
Distance: Maximum 2,500 feet [750 m]
Cable: Belden # 8451 or equivalent.
#22 gauge shielded, twisted pair. (For multiple sensors a multi-conductor
twisted pair, shielded cable should be used).
Signal: 30 VDC - 1.0 Amps.
60 VDC - 0.3 Amps.
125 VAC - 0.5 Amps.
Master Module to Source equipment:
Distance: Maximum 10,000 feet [3,000 m]
Cable: Minimum #22 gauge wire (does not have to be shielded, twisted pair).
Signal: 5 VDC, < 5 µA.
Computer Interface Board:
Output: Dry Contacts NC, open on Alarm.
Rating: 30 VDC - 1.0 Amps.
60 VDC - 0.3 Amps.
125 VAC - 0.5 Amps.