18 Amico Pipeline
SETTING FACTORY DEFAULT
To quickly reset the module (Pressure or Vacuum) to the factory default settings as follows:
- Pressure: High set-point 60 Psi, Low set-point 40 Psi.
- Nitrogen & HP Air: High set-point 195 Psi, Low set-point 140 Psi.
- Vacuum: Low set-point 12 inchHg.
- No Repeat alarm, but set for 30 min..
1. Set switch #8 to the ON position.
2. Turn the power o (wait 5 seconds) then back on.
3. Set switch #8 to the OFF position.
The module is now in the default mode.
SETTING GAS IDENTIFICATION SWITCHES
PLEASE NOTE: DO NOT tamper with switches #1, #2 and #3 on the dip-switch. Tampering with these positions will
result in an error message being displayed (EO2) and will disable the electrical interlock from the gas specic sensor.
Changes to these switches should only be done by properly trained personnel, when circuit
boards have to be changed in the eld.
Switches # 1, #2 and #3 are used for the gas identication of the display module. These will be set at the factory and
should not be tampered with in the eld.