OM006-12 Model 116 Operating Manual Page 15
6 Symbols
The following symbols appear on the audiometer or mains adapter:
Definition: Refer to instruction manual (mandatory).
Definition: Type B applied part an applied part
providing protection against electric shock, particularly
regarding allowable patient leakage current and patient
auxiliary current.
The applied parts are the left & right earphones, patient
response switch and the associated cables.
Definition: The output from the mains AC adapter is
Direct Current.
Definition: Class II equipment equipment in which
protection against electric shock does not rely on basic
insulation only, but in which additional safety precautions
such as double insulation or reinforced insulation are
provided, there being no provision for protective earthing
or reliance upon installation conditions.
7 Technical Information
Audiometer
Audiometer type: Type 4 (IEC 60645-1:2001)
Type 4 (ANSI S3.6:2004)
Battery function
Battery voltage range: 4.0 to 6.0V
Low battery warning: Approx 4.4V
Expected battery life: 6 to 8 hours use from alkaline batteries
Frequency Modulation
Carrier frequencies: 125Hz to 8kHz as per pure tones
Modulation waveform: Sinusoidal
Rising and falling symmetry: Symmetrical on linear frequency scale