Page 16 OM006-12 Model 116 Operating Manual
Modulating frequency: 15.625Hz
Frequency deviation: +/-10%
Transducers
Types and reference levels: DD45: ISO 389-1, Table 2
Static headband force: Headphones: 4.5N
Sound attenuation characteristics: ISO8253-1, Table 3
Earphone Sound Attenuation Characteristics
Frequency, Hz
125
250
500
1000
2000
4000
8000
Attenuation, dB
2
5
7
15
25
31
23
Environmental
Operating temperature: +15
o
C to +35
o
C
Operating humidity: 30% to 90% (non-condensing)
Atmospheric pressure: 700 hPa to 1060 hPa
Input / Output
Power input: 2.5mm barrel-type socket.
Patient response input: 6.3mm Jack socket
Left & Right outputs: 6.3mm Jack socket
USB: Type B socket
Printer: RJ12 socket (6-way)
Maximum voltage at any output: 12V peak
8 Routine Maintenance
8.1 Audiometer maintenance
The Model 116 audiometer is a precision instrument. Handle it carefully in
order to ensure its continued accuracy and service. When cleaning the
instrument, first disconnect it from the mains supply. Use a soft cloth and
mild detergent to clean the instrument panel when required. Refer to ISO
8253-1 for additional guidance.