EasyManua.ls Logo

Amprobe DMIII - 14 Technical Specifications; Technical Features; Safety Test Functions

Amprobe DMIII
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMPROBE
DMIII MULTITEST
EN-76
14. TECHNICAL SPECIFICATIONS
14.1. TECHNICAL FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 60%.
14.1.1. Safety Test functions
LOW: 200mA CONTINUITY TEST (AUTO, RT+, RT- MODE)
Range []
Resolution []
Accuracy(*)
0.01 9.99
10.0 99.9
0.01
0.1
(2% Reading + 2 digit)
(*) After Test leads calibration
Test Current > 200mA DC per R5 (Test leads included)
Resolution for Test current: 1mA
Open Circuit Voltage 4V V0 24V
INSULATION TEST
Test Voltage [V]
Range [M]
Resolution [M]
Accuracy
50
0.01 9.99
0.01
(2% Reading + 2 digit)
10.0 49.9
0.1
50.0 99.9
0.1
(5% Reading + 2 digit)
100
0.01 9.99
0.01
(2% Reading + 2 digit)
10.0 99.9
0.1
100.0 199.9
0.1
(5% Reading + 2 digit)
250
0.01 9.99
0.01
(2% Reading + 2 digit)
10.0 199.9
0.1
200 249
1
250 499
1
(5% Reading + 2 digit)
500
0.01 9.99
0.01
(2% Reading + 2 digit)
10.0 199.9
0.1
200 499
1
500 999
1
(5% Reading + 2 digit)
1000
0.01 9.99
0.01
(2% Reading + 2 digit)
10.0 199.9
0.1
200 999
1
1000 1999
1
(5% Reading + 2 digit)
Open circuit Test Voltage <1.3 x Nominal Test Voltage
Short Circuit Current <6.0mA with 500V Test Voltage
Nominal Test Current 500V >2.2mA with 230k
other >1mA with 1k*Vnom
FREQUENCY MEASUREMENT
Range [Hz]
Resolution [Hz]
Accuracy
47.0 63.6
0.1
(0.1%Reading+1 digit)
RCD and LOOP function are active only for 50Hz 0,5Hz frequency
PHASE ROTATION : VOLTAGE MEASUREMENT
Range [V]
Resolution [V]
Accuracy
0 460V
1
(3%Reading + 2digit)
GROUND TEST: RESISTANCE MEASUREMENT WITH EARTH RODS
Range RE []
Resolution []
Accuracy
0.01 19.99
20.0 199.9
200 - 1999
0.01
0.1
1
(5% Reading + 3 digit)
Test Current <10mA 77.5Hz
Open circuit Test Voltage <20V RMS
GROUND TEST: RESISTIVITY MEASUREMENT
Range
Resolution
Accuracy
0.60 19.99 m
20.0 199.9m
200 1999m
2.00 99.99km
100.0 125.6km(*)
0.01 m
0.1m
1 m
0.01 km
0.1 km
(5% Reading + 3 digit)
(*) setting distance = 10m
Test Current <10mA 77.5Hz
Open circuit Test Voltage <20V RMS