EasyManua.ls Logo

Analog Devices MAX30001 - Page 14

Analog Devices MAX30001
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.analog.com
Analog Devices
│
14
MAX30001 Evaluation System Evaluates: MAX30001
MAX30002
R-to-R and Pace data is shown in the ECG plot. R-to-R peak detection is indicated by a red circle at the peaks of the
R waves. Pace data appears as teal arrows on the rising and falling slopes of the pace signal. Heart rate information
gathered from R-to-R detection can be found in the Heart Rate (from R-to-R) box below the plotting area. Figure 13
illustrates an ECG waveform with R-to-R and pace detection enabled. The ECG sample rate is 512sps and the BioZ
current generator frequency is 80kHz.
Figure 13. An ECG Waveform with R-to-R and Pace Detection Enabled.

Related product manuals