EasyManua.ls Logo

Anritsu MS2690A - DET;DET?

Anritsu MS2690A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter
Native Device Message Details
4-73
4
Native Device Message Details
DET/DET?
Detection Mode
Function
This command selects the waveform pattern detection mode. When
Couple Time/Frequency Domain is set to Off, each parameter is specified
in accordance with the already-specified frequency axis or time axis
measurement.
Command
DET mode
Query
DET?
Response
mode
Parameter
mode
Detection mode selection
NRM
Simultaneous positive and negative peak
detection
POS
Positive peak detection
NEG
Negative peak detection
SMP
Sample detection
RMS
RMS detection
This following functions are available when MS2830A-016/116 or
MS2840A-016/116 is installed.
QPE
QP detection
CAV
CISPR Average detection
CRMS
RMS Average detection
Details
The set detection mode is applied to all traces.
This command is not available during the Spectrum Emission Mask
measurement or the Spurious Emission measurement.
QPE, CAV, or CRMS cannot be set when the Measure function is set to
On.
QPE, CAV, or CRMS cannot be set when the Gate View function is set to
On.
When QPEak, CAVerage, and CRMS are set, RBW MODE changes from
Normal to CISPR.