EasyManua.ls Logo

Anritsu MS2690A - :CALCulate:ACPower:CARRier[1]:RCFRequency?; [:SENSe]:ACPower:CARRier[1]:COUNt <integer>

Anritsu MS2690A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-370
:CALCulate:ACPower:CARRier[1]:RCFRequency?
Adjacent Channel Power In Band Center Query
Function
This command queries the In Band center frequency for Adjacent
Channel Power measurement.
Refer to
[:SENSe]:ACPower:CARRier[1]:RCFRequency?.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:CARRier[1]:RCFRequency?
[:SENSe]:ACPower:CARRier[1]:COUNt <integer>
Adjacent Channel Power Carrier Number
Function
This command sets the carrier number for Adjacent Channel Power
measurement.
Command
[:SENSe]:ACPower:CARRier[1]:COUNt <integer>
Parameter
<integer>
Carrier number
Range 1 to 12
Resolution 1
Suffix code None
Default 1
Details
This command is not available when ACP Reference is set to the
following:
Span Total
Example of Use
To set the carrier number to 12.
ACP:CARR:COUN 12
Related command
This command has the same function as the following command.
:CALCulate:ACPower:CARRier[1]:COUNt <integer>