2.14 Spurious Emission Measurement Function
2-741
2
SCPI Device Message Details
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP?
Spurious Emission Stop Frequency Query
Function
This command queries the stop frequency of each segment.
Query
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP? <freq_n>
Response
<freq_n>
Parameter
<freq_n>
Stop frequency of Segment n
[MS269xA]
–99.9997 MHz to 6.05 GHz (MS2690A)
–99.9997 MHz to 13.6 GHz (MS2691A)
–99.9997 MHz to 26.6 GHz (MS2692A)
[MS2830A]
–99.9997 MHz to 3.7 GHz (Option 040)
–99.9997 MHz to 6.1 GHz (Option 041)
–99.9997 MHz to 13.6 GHz (Option 043)
–99.9997 MHz to 26.6 GHz (Option 044)
–99.9997 MHz to 43.1 GHz (Option 045)
[MS2840A]
–99.9997 MHz to 3.7 GHz (Option 040)
–99.9997 MHz to 6.1 GHz (Option 041)
–99.9997 MHz to 27 GHz (Option 044)
–99.9997 MHz to 45 GHz (Option 046)
Resolution 2 Hz
Suffix code None. Value is returned in Hz units.
Example of Use
To query the stop frequency of each segment.
SPUR:FREQ:STOP?
>100000,40000000,1100000000,1000000000,1000000000,100000
0000,1000000000,1000000000,1000000000,1000000000,1000000
000,1000000000,1000000000,1000000000,1000000000,10000000
00,1000000000,1000000000,1000000000,1000000000