[(16QAM, FullRB) measurements]
39. Execute ULRMC_RB 25 to set UL RB number to 25.
40. Execute TP_MPR1_UL 25.7 to set TX Power measurement Pass/Fail upper limit to 25.7 dBm.
41. Execute TP_MPR1_LL 16.8 to set TX Power measurement Pass/Fail lower limit to 16.8 dBm.
42. Execute steps 14 to 17.
NOTE 1: The UL RB Position for PartialRB allocation is divided as follows:
When Test Frequency is Low range, Max (#max)
When Test Frequency is Mid range, Min (#0) and Max (#max)
When Test Frequency is High range, Min (#0)
NOTE 2: There is no need to set separately because the Pass/Fail evaluation value for Spectrum Emission
Mask measurement changes in accordance with the additionalSpectrumEmission setting.
NOTE 3: The usual Pass/Fail evaluation value is set to the Band 1 default value described in TS36.521-1.
Since the evaluation value differs according to the Band, set the evaluation value described in
TS36.521-1 Table 6.2.4.5-1 at:
TP_MPR1_LL
TP_MPR1_UL
For the Pass/Fail evaluation values, refer to Chapter 3.7.4 Test Parameter Limit in the
operation manual.
Additional Maximum Power Reduction (A-MPR) for HPUE (6.2.4_1) 3.1.34.
Because there are no test parameters supporting Additional Maximum Power Reduction for HPUE test, select the
basic parameter (TX1 - Max. Power (QPSK/FullRB)) and set parameters and the pass/fail evaluation limits for each
test condition/requirement.
This chapter describes UL measurement examples where (Modulation, RB) are (QPSK, PartialRB), (QPSK, FullRB),
(16QAM, PartialRB) or (16QAM, FullRB) when additionalSpectrumEmission is NS_06 and Test Frequency is Mid range.
1. Execute BAND 14 to set Operating Band to 2.
2. Execute PWR_AVG 20 to set the average count of power measurement to 20 times.
3. Execute SEM_AVG 20 to set the average count of Spectrum Emission Mask to 20 times.
4. Execute TESTPRM TX_MAXPWR_Q_F to set Test parameter to TX1 - Max. Power (QPSK/FullRB).
5. Execute ALLMEASITEMS_OFF to set fundamental measurement items to OFF at one time.
6. Execute PWR_MEAS ON to set Power measurement to ON.
7. Execute SEM_MEAS ON to set Spectrum Emission Mask measurement to ON.
8. Execute SIB2_NS NS_06 to set additionalSpectrumEmission to NS_06.
[(QPSK, PartialRB) measurements]
9. Execute ULRMC_MOD QPSK to set UL RMC modulation to QPSK.
10. Execute ULRMC_RB 8 to set UL RB number to 8.
11. Execute ULRB_POS MIN to set UL RB Position to Min (#0).
12. Execute TP_MPR1_UL 33.7 to set TX Power measurement Pass/Fail upper limit to 33.7 dBm.
13. Execute TP_MPR1_LL 26.3 to set TX Power measurement Pass/Fail lower limit to 26.3 dBm.
14. Execute SWP to measure the power.
15. Execute POWER? AVG to read the TX Power measurement result.
16. Execute POWERPASS? to check the measurement result is PASS.
17. Execute SEMPASS? to check the SEM result is PASS.
18. Execute ULRB_POS MAX to set UL RB Position to Max (#max).
19. Execute step 14 to 17.