Location Registration and Packet Connection 5.3.5.
1. Execute CALLSO to clear call processing.
2. Execute CALLSTAT? to confirm the call processing status is 1 (= Idle).
3. Turn on the UE power.
4. Execute CALLSTAT? to confirm the call processing status is 6 (= Connected).
Repeat Polling query response when the checked status is not 6 (= Connected).
5. Execute TPUT_SAMPLE 2000 to set the number of Throughput measurement samples to 2000.
6. Execute SWP to perform the Throughput measurement.
7. Execute TPUT? PER to read the Throughput measurement result (%).
At an error, the DL transmission condition must be optimized by changing the output level or RMC setting of
each CC by referring to Chapter 5.3.4.
Fig. 10.6-1 Throughput Measurement Result Screen of DL CA (Fundamental Measurement)