227
33. Execute LINKEPSID 5 to set Call Processing Parameter - Packet - Linked EPS Bearer Identity to 5. (NOTE
1)
34. Execute TFTIPV4 192,168,20,100 to set Call Processing Parameter - Packet - TFT Remote IPv4 Address to
192.168.20.100.
35. Execute TPUT_MEAS ON to set Fundamental Measurement Parameter - Throughput Measurement to On.
NOTE 1:
Some UEs may request establishment of the second Default EPS Bearer by sending a PDN Connectivity
Request message after confirming the Connected state. To verify IP data communication with this UE and
communicate with the second EPS Bearer, execute following procedure instead of procedure No. 33.
33. Execute LINKEPSID 6 to set Call Processing Parameter - Packet - Linked EPS Bearer Identity to 6.
The Dedicated EPS Bearer will be linked to the second Default EPS Bearer by this setting. The SCC using the
Dedicated EPS Bearer as the IP data path will communicate with the second Default EPS Bearer IP address
(Call Processing Parameter - Client IP Address2).
[Example of IP Data Path when Linked EPS Bearer Identity set to 6]
SCC can communicate with the second EPS Bearer IP address (Call Processing Parameter - Client IP Address2).
Figure 5.3.4-1 IP Data Path (Linked EPS Bearer Identity = 6)