CAUTION!
Turn the device off prior to exchanging the batteries.
Make sure the sample area is empty and dry before exchanging the batteries to avoid spillage of liquids.
Only open the device in a clean and dry environment.
The inside of the device does not require any cleaning. Do not use liquid cleaning agents on interior parts of the device.
Do not open the compartment containing the optical and electronical components of the device as this will void the warranty.
STEPS TO EXCHANGE THE BATTERIES
Unscrew the bottom cover of the device using an Allen key (2 mm; 5/64 in)
Remove the empty batteries.
Insert the fresh batteries with the correct orientation!
Reattach the bottom cover and tighten the screws rmly.
NOTICE
Ensure that the screws are tightened properly so that no liquid can get inside. Do not overtighten the screws. Using excessive
force may permanently damage your device. After replacing the batteries, be sure to only use the original screws to x the
battery compartment cover. Consider the polarity of the batteries when you insert them in the battery compartment. Insert
only equally charged batteries of the same type.
NOTICE
When replacing the batteries, check if the O-ring between the bottom and the top cover is not damaged and in place. Make