PROGRAMMING C
NTA47262-000A6 17 / 55
c) "Low Inductive Resistance and Automatic Triggering" Measurement Mode
This mode is intended only for measuring resistance levels without a time constant. The use of the dagger
prod connector detailed in the accessories section is recommended.
From the standby position, to start a cycle:
Press to start the cycle.
Connect the wires to the resistor. The unit waits until it detects that the currents and voltage wires are
connected correctly.
Residual voltage (U0) measurement across the resistor terminals.
The measurement current (I) is established, the voltage across the resistor terminals (U1) I measured and
the measurement R = (U1-U0)/I is displayed.
To start a new measurement cycle: Release at least one connection, then reconnect it.
To stop the cycle, press .
REMARKS:
If the range is exceeded, the unit displays Err 07.
Any change of range will stop the cycle and the unit returns to the standby position.
It is possible to store and compensate measurements while the cycle is running.