Normes applicables F
NTA47262-000A6 47 / 55
F. Specifications
F.1 Applicable Standards
F.2
Safety Class F.2.1
In accordance with European standard EN 61010-1.
Category II, pollution 2.
Rated voltage: 60 V.
.
Note 1: The safety instructions relating to the unit are described in Section 1.
EMC Conformity F.2.2
According to European standard EN 61326, standard environment, non-continuous criterion.
The unit’s performance levels meet the demands of the following norms:
Conducted and radiated emission:
EN 55022, classe B EN 61000-3-2
EN 61000-3-3
Immunity :
EN 61000-4-2 electrostatic discharges
EN 61000-4-3 radiated fields
EN 61000-4-5 shock waves
EN 61000-4-6 conducted disturbances
EN 61000-4-11 voltage drops
EN 61000-4-4 bursts
Ambient Conditions F.2.3
In accordance with IEC Publication 359 (national standards NF C 42-600, DIN 43745): operating
category I.
Reference range: 23°C 5°C, relative humidity: 45% to 75%.
Normal operating range: 0°C to + 50°C, relative humidity: 20% to 80% non-condensing.
Operating range limits: - 10°C to + 55°C, relative humidity: 10% to 80% non-condensing.
Storage and transport range: - 40°C to + 60°C (- 15°C to + 50°C with battery charged).
Mechanical Conditions F.2.4
Sealing according to IEC publication 529 (national standard NF C 20-010):
IP53 = case open.
IP64 = case closed.
Impacts: according to European standard EN 61010-1.
Measurement under Reference Conditions
F.2.5
According to IEC publication 485 (national standards NF C 42-630 and DIN 43751).