PROGRAMMING C
NTA47262-000A6 21 / 55
C.3 List of Error Messages
Error number
Comments
Err 01
Battery too low.
Err 02
Internal problem.
Err 03
Cannot measure battery charge.
Err 04
Cannot measure the temperature.
Err 05
Excessive current source heating.
Allow it to cool.
Err 06
Measurement current level not established.
Err 07
Measurement out of range.
Err 08
Internal problem.
Err 09
Measurement cycle stopped.
Err 10
Temperature sensor incorrectly
connected or missing.
Err 11
"Current" wires incorrectly connected.
Err 12
"Voltage" wires incorrectly connected or
measured resistance level too high.
Err 13
Residual voltage level too high.
Err 21
Adjustment value out of range.
Err 22
Measured value out of range.
Err 23
Entry out of range.
Err 24
Cannot write to the protected memory.
Err 25
Cannot read from the protected memory.
Err 26
Memory full.
Err 27
Memory empty, no data available.
Err 28
Memory check problem.
Err 29
Incorrect "object" or "test" number.
WARNING: If error messages 2, 3, 4 and 8 are displayed, the unit must be shut down and returned for
servicing.