COMMUNICATION
30 / 55 NTA47262-000A6
OM 17
Command ErrorsD.1.1
If the unit does not understand the command received or if its arguments are incorrect, it sets the ERR_NO
error code and does not reply even if a reply is called for.
The unit places the successive error codes in a four-element FIFO type memory (where the first error
recorded is the first error removed).
The two ERR_NO? and ERR? commands are used to obtain the error codes in the order in which they
appeared. The CL_ERR command clears the contents of the FIFO.
List of error codes:
ERROR
o.
DESCRIPTION
MEANING
0
NONE ERROR
No error.
1
UNKNOWN HEADER
Unknown command header.
2
ARG. TOO LONG
Argument too long for the unit.
3
WRONG ARG. NB.
The argument number is the wrong one for the command received.
4
OVERLIMIT ARG.
One of the arguments is outside of the limits allowed for this type of
argument.
5
UNKNOWN MNEMONIC
The argument is an unknown mnemonic.
6
WRONG SUFFIX
Suffix error.
7
WRONG ARG. TYPE
One of the arguments is the wrong type for the command received.
8
LOCAL
The unit is in "keyboard active" mode when the command required
that it be set to REMOTE mode.
9
WRONG ERROR NO
Wrong error number.
10
CALIBRATION ERROR
Error during calibrating.
11
WRONG ARG.
One of the arguments is incompatible with the command received.
12
NOSTORAGE MEMORY
There is no measurement file in memory.
13
READ MEMORY
Cannot read memory.
14
WRITE MEMORY
Cannot write memory.
15
LIMIT CONF.
The limit configuration is incorrect.
16
CORR. CONF.
The correction configuration is incorrect.
17
WRONG CAL.
Wrong calibration.
18
IMPOSSIBLE ADJUST
Cannot perform adjustment.