RS 232 CONNECTIONS D
NTA47262-000A6 33 / 55
D.4 Programming Commands
CFG <ModeMes>, <Range>
Description: Programming the general measurement configuration parameters.
The unit must be in REMOTE mode
Argument :
<ModeMes>
measurement mode: inductive = SELF, low inductive = ASELF, auto = AUTO
<
Range
> MOHM5, MOHM25, MOHM250, MOHM2500, OHM25, OHM250, OHM2500.
Reply : None.
Example : CFG SELF, MOHM250
CFG?
Description: Reading the general measurement configuration parameters.
Argument : None.
Reply :
<ModeMes>, <Range>
(as described above)
Example : CFG?
SELF, MOHM250
LIMIT <NumSeuil>, <En_Hors>[, <ValSeuil>[, <Unité>[, <Sens>[, <Buzzer>]]]]
Description: Programming a limit value.
The unit must be in REMOTE mode
Arguments :
<NumSeuil>
1 or 2
<En_Hors>
OFF or ON
<ValSeuil>
Threshold value
<Unité>
Unit: OHM, MOHM
<Sens> LO or HI
<Buzzer> BUZ_NONE (no buzzer) , BUZ_LO (low), BUZ_HI (high).
Arguments omitted from the command retain their current value
Reply : None.
Examples : LIMIT 1, ON, 0.246, OHM, HI, BUZ_LO
LIMIT 1, OFF
LIMIT 1, ON
LIMIT? <NumSeuil>
Description : Limit (threshold) programming request
Argument : <NumSeuil> NumSeuil = 1 or NumSeuil = 2
Reply :
<En_Hors>, <ValSeuil>, <Unité>, <Sens>, <Buzzer>
(as described above).
Example : LIMIT? 1
OFF, 0.246, OHM, HI, BUZ_LO
TCOMPENSATION < En_Hors >[, <Tréférence>, <Unité>]
Description: Programming temperature compensation.
The unit must be in REMOTE mode.
Argument :
<En_Hors>
ON or OFF
<TReference>
Temperature that the measurement is returned to,
expressed in the unit <
Unité
>.
<Unité> Temperature display unit: CEL (°C) or FAR (°F).
Reply : None.
Example : TCOMPENSATION ON, 23, CEL