TCOMPENSATION?
Description: Temperature compensation parameter request.
Argument : None.
Reply :
<EnService>
ON or OFF
<TRéférence>
Temperature that the value is brought back to.
<Unité
> Measurement unit, CEL (°C) or FAR (°F).
Example : TCOMPENSATION?
ON, 23, CEL
METAL <TypeMetal>[, <Alpha>]
Description: Programming the type of metal and the alpha for the OTHER metal.
The unit must be in REMOTE mode
Argument :
<TypeMetal>
CU, AL, OTHER ->Type of metal for temperature correction purposes
<Alpha> Temperature coefficient of the "OTHER" metal, as 10
-3
/°C.
e.g. silver = 3.85 x 10
-3
/°C, enter 3.85
Remark: the Alpha value is only used for "OTHER METAL". If it is entered with the command,
it will be saved for "OTHER METAL" but not taken into account for CU and AL.
Reply : None.
Example : METAL CU Select copper as the metal
METAL OTHER 3.85 Select another metal and its alpha temperature coefficient. For silver,
alpha = 3.85
METAL?
Description: Returns the parameters for the selected metal.
Argument : None
Reply :
<TypeMetal>
CU, AL, OTHER ->Type of metal for measurement purposes
<Alpha> Temperature coefficient of the OTHER METAL, as 10
-5
/°C.
Example : METAL?
CU, 3.85 Copper followed by the alpha temperature coefficient of the OTHER
METAL.
METAL ?
OTHER,3.85 OTHER METAL followed by its alpha value.
TAMBIANT <InfoMesure>[, <Valeur>, <Unité>]
Description: Programs the ambient temperature parameters.
The unit must be in REMOTE mode
Argument : <
InfoMesure>
MEAS (by the Pt100 probe) or ENTRY (by the user).
<
Valeur>
Temperature value entered by the user.
<Unité
> Temperature unit: CEL (°C) or FAR (°F).
Reply : None.
Examples : TAMBIANT MEAS Temperature measured by the Pt100 probe.
TAMBIANT MEAS, 25.1, CEL Ambient temperature measured by the Pt100 probe.
Temperature entered, stored, but not taken into account.
TAMBIANT ENTRY, 25.1, CEL Ambient temperature entered.
TAMBIANT?
Description: Returns the ambient temperature parameters.
Argument : None
Reply :
<InfoMesure>
MEAS, measured by the Pt100 or ENTRY, entered by the user.
<Valeur>
Temperature value entered by the user or the last value measured.
<Unité >
Temperature unit CEL (°C) or FAR (°F).
Examples : TAMBIANT?
MEAS, 24.6, CELTemperature measured by the Pt100 probe.
TAMBIANT?
ENTRY, 25.1, CEL Temperature entered.