RS 232 CONNECTIONS D
NTA47262-000A6 37 / 55
TEST? <NoObjet>, <PositionTest>
Description: Reading an object test.
The unit must be in REMOTE mode
Argument :
<NoObjet> Object
number (1 to 99).
<PositionTest>
Position in the object of the desired test (1 to 99).
Reply : #218
<TEST binary structure 18 bytes
> <LF>
The test's 16 data bytes are described by the following structure:
STRUCTURE
BIT
COMMENTS
MSB
NumTest
8
Test number
TypeMes
2
Measurement mode: 1=low inductive, 2=inductive, 3=auto
TypeMetal
2
Cu=1, Al=2, Other=3 ->Type of metal for measurement
Cal
3
Calibration No., from 1 (=5 mΩ) to 7 (=2500 Ω).
InfoPt100
1
Tamb: Measured by Pt100 = 1, meas. entered by user = 0
**For threshold 1**
SensHaut1
1
Threshold1: Direction down = 0, up = 1
Actif1
1
Threshold1: Threshold inhibited = 0, active = 1
UnitOhm1
1
Threshold1: Unit mΩ = 0, Ω = 1
Cpav1
3
Threshold1 : Number of digits after the decimal point (max. 4)
Depasse1
1
Threshold exceeded = 1, if not = 0.
InfoUnitDeg
1
Temperature display unit Cel = 0, Far =1
**For threshold 2**
SensHaut2
1
Threshold2: Direction down = 0, up = 1
Actif2
1
Threshold2: Threshold inhibited = 0, active = 1
UnitOhm2
1
Threshold2: Unit mΩ = 0, Ω = 1
Cpav2
3
Threshold2: Number of digits after the decimal point (max. 4)
Depasse2
1
Threshold exceeded = 1, if not = 0.
Correction
1
If correction = 1; if not = 0
ValSeuil1
16
Value threshold 1
ValSeuil2
16
Value threshold 2
Tref
16
Temperature that the value is returned to
(unit = one hundredth of a °C -> 2000 20.0°C)
Tamb
16
Ambient temperature when entered manually
Alpha
16
OTHER metal alpha temperature coefficient, in 10
-3
/°C
Mesure
16
Measurement made (not compensated)
MesureTref
16
Measurement compensated
LSB
CLR_OBJECT <Objet>
Description: Clears all of the
Tests
for an
Object
, and clears the
Object
.
The unit must be in REMOTE mode
Argument : <
Objet>
number (1 to 99) of the
Object
to clear.
Reply : None.
Example : CLR_OBJET 12
CLR_ALL_OBJECTS
Description: Clears all
Objects
.
The unit must be in REMOTE mode
Argument : None.
Reply : None.