lightly wipe the catholic of gold plane (along the arc of the gold surface), and clean
the surface of cathode with soft tissue), and wash electrode with distilled water and
throw off the water on it. Add some electrolyte into the membrane cap, re-assembly
and screw tightly. Then conduct the full scale and zero oxygen calibration according
to 4.2. and 4.7.
4.8 Salinity Calibration
The meter has automatic salinity compensation, so salinity calibration is needed when
replacing a new electrode or the electrode has not been used in a long time. In
general, there is no need to do salinity calibration. Salinity has already been calibrated
before leaving factory. For the procedure of salinity calibration, please refer to 4.5.3 in
P2.
4.9 Barometric pressure setting
The meter has the function of manual Barometric pressure compensation. It is
suggested to reset the value according to the standard Barometric pressure table
when there is a big change in Barometric pressure in order to ensure the accuracy of
Barometric pressure compensation (Please see the Chart Ⅲ and Ⅳ. For the
procedure of Barometric pressure calibration, please see 4.5.4.).