For Crest Factor of 1.4 to 2.0, add 1.0% to AC accura-
cy.
For Crest Factor of 2.0 to 2.5, add 2.5% to AC accura-
cy.
For Crest Factor of 2.5 to 3.0, add 4.0% to AC accura-
cy.
Frequency Counter
Minimum Sensed Frequency : 5Hz
Range Resolution Accuracy
400.0Hz
4.000kHz
40.00kHz
100.0kHz
0.1Hz
1Hz
10Hz
100Hz
1d (3¾-digit mode)
5d (4¾-digit mode)
Frequency Counter Sensitivity
Function Range
Sensitivity (Peak to Peak)
5 to 10k Hz 10k to 100k Hz
mV
V
mA
A
40.00mV
400.0mV
4.000V
40.00V
400.0V
1000V
40.00mA
400.mA
4.000A
10.00A
10mV
40mV
0.4V
4V
40V
400V
10mA
40mA
1A
4A
Unspecified
100mV
1V
10V
Unspecified
Unspecified
501 / 502
EN