Simpool pH model INSTRUCTION AND MAINTENANCE MANUAL
Programmable digital peristaltic pump ENGLISH
ADSP7000407 rev. 1.2 27/03/2015 55/313
Fig. 14- pH Probe Calibration Menu
3.3.4 SETTING THE TEMPERATURE
The temperature setting allows to correct the pH value read.
From the Programming Menu, by accessing it from the °C entry you can choose between the “noA” manual setting
and the ”Aut” automatic temperature reading from the PT100 probe or OFF.
If you want to correct the pH value in relation to the thermal variation, in manual mode just set a temperature equal
to 25°C or select OFF.
The default setting is the manual mode, with a temperature of
25°C (pH correction not enabled).
If you have not selected OFF among the options, the
following will be displayed by pressing Enter:
Temperature value measured from 0.0 to 99.9°C if
the mode is Aut: if the probe is not connected, the
display will show “t - -“;
Set temperature value, from 0 to 99.9 °C if the mode
is noA;
To go back to the Programming Menu, press Enter again.
FIG. 15- TEMPERATURE