ARGUS service
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 24 / 48
4.4.5. ARGUS service - Toolbox - Pump history
Each registered event has his own date & time stamp. An event is
registered on each pump status change. Please refer to the com-
plete list mentioned in chapter 4.4.6 below.
4.4.6. History messages
Possible messages appearing in the description of each history event:
No information available Pump start Pump start in remote mode
Battery defective Pump stop (KVO) Rate change during remote mode
Battery low prealarm Pump off Pump off in remote mode
Battery low, pump stop Pump on Enter setup mode
Airbubble, pump stop Transport off Enter PC configuration mode
Door open, pump stop Transport on Exit setup or PC configuration mode
No drops, pump stop Data lock off Downstream occlusion, pump stop
Not enough drops, pump stop Data lock on Upstream occlusion, pump stop
Too many drops, pump stop Rate change Timer alarm, pump stop (KVO)
Neonatology mode set in PrL Inf-Set change PC communication timeout reached
PrLimit change Infusm cleared Logon in PC configuration mode
Bolus total reached Bolus start Logoff in PC configuration mode
Pump has detected failure Bolus stop Pump start, ext. changed parameters
Total volume reached, pump stop External power on Any defaults written in EEPROM area
Total (VTBI) change External power off CRC error in PC configuration module
4.4.7. History printout example