REPLACEMENT OF PARTS
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 42 / 48
7. Use the short screw at these
places (torque 1.1 Nm)!
7. Use the long screw at these
places (torque 1.1 Nm)!
Mind the torque for the screws listed below:
What: Where Torque in Nm
Case Backplane into front 1.10
Pump unit Into front 1.00
Combination clamp Into clamp holder 1.50
Pressure sensor Into front 0.60
Stop flow Into front 0.60
Air detector Into front 0.60
Sensor board Into pump unit 0.40
Display board Into front 0.50
Transformer Into Backplane 1.00
Clamp holder Into Backplane 0.50
Mains plug Into Backplane 0.50
Edge board Into clamp holder 0.50