REPLACEMENT OF PARTS
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 30 / 38
5. Reconnect the back
side with the front side
(battery, power &
docking interface)!
6. Make sure that both
cables are correctly
placed in the notch be-
fore the main board is
inserted!
7. Use the short screw at these
places (torque 1.1 Nm)!
7. Use the long screw at these
places (torque 1.1 Nm)!