33
Miscellaneous
EMC
First Step Select
MRS uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause interference in nearby electronic equipment. The First Step Select
MRS was tested to the following standards.
EN 60601-1-2: 2007
Conducted Emissions: Nov. 21, 2008 (Group 1, Class B)
Radiated Emissions: Dec. 17, 2008 (Group 1, Class B)
EN 61000-4-2: Dec. 3, 2008 (6/8 kV)
EN 61000-4-3: Dec. 1, 2008 (3 V/m)
EN 61000-4-4: Dec. 18, 2008 (2 kV)
EN 61000-4-5: Dec. 1, 2008 (1 kV)
EN 61000-4-6: Dec. 24, 2008 (3 Vrms)
EN 61000-4-8: Dec. 25, 2008 (3 A/m)
EN 61000-4-11: Dec. 21, 2008 (>95%-0.5p, 30%-25p,>95%-250p)
EN 61000-3-2: Dec. 21, 2008 (Class A)
EN 61000-3-3: Dec. 24, 2008
Classification
Classication in accordance with UL 60601-1, CSA C22.2 No 601.1 and IEC. 60601-1
Class II (UL 60601-1)
Type B
No Sterilization
Not for Use with Flammable Anesthetic Mixture with Air, Oxygen or Nitrous Oxide.
Continuous Operation
Ordinary Equipment
Environmental Storage / Transit Conditions
Temperature Range: -20˚ to 60˚ C (-4˚ to 140˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa
Environmental Operating Conditions
Temperature Range: 10˚ to 40˚ C (50˚ to 104˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa