Page 12 · 5300 Series TECSource User’s Manual
Measurements
The second line of the display contains the measurements appropriate for the
control mode. In T Mode, the actual temperature, current, and voltage can be
displayed, depending on the display mode. In R Mode, the actual sensor value
is displayed instead of the actual temperature. In Ite Mode, the actual
temperature is displayed.
No Set Point Error
Because 5300 Series TECSource uses a digital PID loop, it will always stabilize
at the set point, and you will never have a temperature offset, which is different
that most other temperature controllers.
Using Limits
The TECSource supports three different types of limits: temperature, sensor,
and current. There is both a temperature high and temperature low limit, and if
the actual temperature exceeds either of these limits, the output will be shut off.
Likewise, both high and low limits exist for the sensor (R limits), and the output
will be shut off if the sensor measurement exceeds either of these limits.