Page 38 · 5300 Series TECSource User’s Manual
Error Messages
Error
Code
Description Cause
E-100 General Error The error code is non-specific, and is
generally used when no other error code is
suitable.
E-102 Message too long The message is too long to process
(USB/Serial only).
E-104 Type not allowed The RADix type was invalid
E-123 Path not found The message used an invalid path
command (USB/Serial only).
E-124 Data mismatch The message contained data that did not
match the expected format (USB/Serial
only).
E-126 Too few or too many
elements
The command requires more or less than
the number of parameters actually supplied.
E-127 Change not allowed An attempt was made to change a
parameter that cannot be changed, or is
currently read-only.
E-201 Data out of range The message attempted to set a value that
was outside the allowable range (USB/Serial
only).
E-202 Invalid data type When trying to parse the message, the data
was in an invalid format (USB/Serial only).
E-204 Suffix not valid An invalid number base suffix (radix) was
encountered when parsing a number
(USB/Serial only).
E-402 Sensor open, output
turned off
A sensor open circuit was detected and the
output was turned off.
E-403 Module open, output
turned off
A Peltier module open circuit was detected
and the output was turned off.
E-404 I limit, output turned
off
A current limit was detected and the output
was turned off.
E-405 V limit, output turned
off
A voltage limit was detected and the output
was turned off.
E-406 Thermistor resistance
limit, output turned off
The thermistor resistance limit (high or low)
was exceeded and the output was turned
off.
E-407 Temperature limit,
output turned off
The temperature limit (high or low) was
exceeded and the output was turned off.
E-409 Sensor changed,
output turned off
The sensor type was changed, causing the
output to be turned off.
E-410 Temperature was out
of tolerance, output
turned off
The temperature went out of tolerance and
the output was turned off.