EasyManua.ls Logo

Arroyo Instruments TECSource 5305 - Working With RTDs; RTD Resistance-Temperature Equations

Arroyo Instruments TECSource 5305
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 24 · 5300 Series TECSource User’s Manual
Working With RTDs
Like thermistors, RTDs also function by converting temperature into resistance,
but unlike thermistors, RTDs increase in resistance as temperature increases.
RTDs are also a fairly linear device, meaning they can be used across a much
broader temperature control range.
According to IEC751, the resistance/temperature relationship is determined
using one of two equations, dependent on the temperature or resistance value
being measured. For resistances above the R
0
value (resistance at 0°C, typically
100) of the RTD, the following equation is used:
)1(
2
0
BTATRR
Below R
0
, an additional term is added to the equation:
])100(1[
32
0
TTCBTATRR
In both of these equations, R
0
is the resistance of the RTD at 0°C, and A, B, and
C are the coefficients as defined by IEC751, through regression analysis, or by
using the Callendar-van Dusen method. By default, the TECSource uses the
Laboratory standard coefficients, which are for a 0.003926/ /°C curve (A =
3.9848x10
-3
, B = -0.58700x10
-6
, C = 4.0000x10
-12
, and R
0
= 100).
The TECSource automatically determines which equation to use based on the
conditions at the time of the measurement.

Related product manuals