Prepare Alternative Solution:
Note: Record weights, in grams, of MVS Stock Solution and solvent during
sample preparation.
Place a clean, amber bottle with cap on the balance pan and press tare.
Add the amount of MVS Stock Solution determined using the Alternative
Solution Helper instructions above. Replace the cap on the bottle.
Record the weight once the balance reaches equilibrium.
Remove the cap, place it on the balance pan and press tare.
Add the appropriate amount of solvent determined using the Alternative
Solution Helper.
Replace the cap on the bottle and record the weight of solvent when the
balance has reached equilibrium.
Mix each Alternative Solution by inversion approximately 20 times.
Label the bottle with the Stock and Solvent weights, the Solvent Density and
the date.
Allow the solution to equilibrate to room temperature before using in any tests.
Add the Alternative Solution to Data Manager.
Alternative Solution Volume Range per Plate Type
Plate Type
Supported
Volume Range
(
µL)
MVS Stock
Solution
96 Well Standard
Profile
0.4 – 9.9
10 – 49.9
1
2
384 Well Standard
Profile
0.1 – 2.49
2.5 – 9.9
1
2
384 Round Well
Low Volume
0.05 – 1.49
1.5 – 3.99
1
2
384 Well Low
Profile
0.04 – 1.49
1.5 – 3.99
1
2
15
Alternative Solution Preparation
(continued)