After all required microtiter plate readings have been collected, Data
Manager will display a Test Report (see screen image on page 21). Reports
may also be automatically exported by checking the box in the Options
menu.
The Test Report can be exported in HTML or XML format and may be
imported to a spreadsheet program for further analysis.
The yellow or orange colored cells represent dispensed volumes or dilution
ratios that exceed the user-defined limits for Relative Inaccuracy and/or
Coecient of Variation.
If the Heat Map was enabled in the Plate Layout, the Test Report will
display colored data points representing the variation in volumes within
that Layout Group.
If the Security System is active, Data Manager will prompt for a password
once all scans are completed. This process attaches an electronic signature
to the end of the Test Report and logs the action in the Audit Trail.
The electronic signature should be considered the equivalent of a dated,
handwritten signature applied to the test result.
Comments that are added to the report are saved, but any changes to the
way data is viewed, such as Change Limits, will not be saved.
Test Reports