2. Intended use
PHG 300A / 301A User Manual Page 6 of 40
2. Intended use
Multifunctional unit PhysioGo 300A / 301A is an active, non-invasive therapeutic device, intended for carrying
our treatment procedures using:
bipolar (bidirectional) and unipolar (unidirectional) low frequency currents,
bipolar (bidirectional) medium frequency currents and unipolar (unidirectional) medium frequency currents
modulated by low frequency waveforms,
using ultrasound therapy and phonophoresis,
combination method of current and ultrasounds.
Model 300A Non battery unit
Model 301A Battery unit
The unit is equipped with three fully independent treatment channels. The list of available therapies in particular
channels is shown in the table below.
Channel Therapy
1
Single circuit electrotherapy – A
Dual circuit electrotherapy A+B
2
Single circuit electrotherapy – B
Combined therapy ultrasound therapy and electrotherapy
3
Ultrasound therapy
Detailed information about available configurations is presented further in this manual.
The unit possesses the base of preset treatment procedures along with therapeutic encyclopedia, which
significantly increases comfort of operation.
There is also a possibility to create own user-defined:
programs for all therapies,
sequences for electrotherapy.
In the range of ultrasound therapy, PhysioGo unit may cooperate with the following types of ultrasound heads:
dedicated to standard ultrasound therapy with the effective radiation area 4 cm
2
, which generates the
ultrasound wave with 1 MHz or 3,5 MHz frequency,
dedicated to standard ultrasound therapy with the effective radiation area 1 cm
2
, which generates the
ultrasound wave with 1 MHz or 3,5 MHz frequency.
PhysioGo unit model 300A / 301A enables the simultaneous connection of two ultrasound heads. In manual
mode, selection of ultrasound head type is done by the field marked with a symbol . In the case of preset
treatment programs and user-defined programs, when you start the procedure the head type selection window
will appear.
The unit may perform treatments by:
interferential currents dynamic and isoplanar,
one-channel sine wave current (AMF),
Kotz’ – Russian stimulation,