Connection of circulating constant temperature bath
6
WARNING
ەFor connecting a circulating constant
temperature bath with the refractometer,
make sure to use the elicon tube (made
of vinyl chloride).
Moreover, tightly fasten each joint part
between the nozzle and tube with the
tie band.
ە
The temperature of hot water to be
supplied from a circulating constant
temperature bath to this refractometer
is limited to 120Υ, and the flow rate of
the circulating constant temperature
bath should be 10 L/min. at maximum
through a nozzle of 6mm caliber.
To o much discharge not only increases
water pressure but also may cause
disconnection between the nozzle and
tube or may overload the water channel
inside the refractometer.
ە
If the tube is damaged, replace it with
new one as immediately as possible.
If the damaged tube is continuously used,
the operator may scald himself with hot
circulating water.
ەPay careful attention to hot circulating
water.
1. Precautions for use