D620.388.00_QS_UK_01
22 / 42
PNEUMATIC CONNECTORS
Pneumatic connectors used to connect the part under test are located on the back panel
of the device.
Pneumatic supply
The pneumatic supply has to meet specific requirements recommended by ATEQ. Refer to
Good practices and safety instructions section.
A specific filter may be necessary.
The air is supplied via the filter located on the back panel of the device.
Metallic model air filter
The metallic model filter is used for 1MPa
range.
The maximum pressure admissible is 1.2
MPa.
Plastic model air filter
The plastic model filter is used for 0.5
MPa range (direct and indirect modes) or
2MPa range (indirect mode only).
The maximum pressure admissible is 690
kPa.